Instrument Network and Security
Refer to Illumina Instrument Control Computer Security and Networking for further information on instrument security and network connections. The following sections include security and networking information specific to the NovaSeq 6000Dx .
Illumina does not install or provide technical support for network connections. Review network maintenance activities for potential compatibility risks with the NovaSeq 6000Dx Instrument.
Use the following guidelines to install and configure a network connection:
| • | Use a dedicated 1 gigabit connection between the instrument and data management system. Make this connection directly or through a network switch. |
| • | Required bandwidth for a connection is as follows. |
| – | 200 Mb/s/instrument for internal network uploads. |
| – | 200 Mb/s/instrument for BaseSpace Sequence Hub network uploads. |
| – | 5 Mb/s/instrument for Instrument Operational Data uploads. |
| • | Switches must be managed. |
| • | Network equipment such as switches must have a minimum capacity of 1 gigabit per second. |
| • | Calculate the total capacity of the workload on each network switch. The number of connected instruments and ancillary equipment, such as a printer, can impact capacity. |
| • | If possible, isolate sequencing traffic from other network traffic. |
| • | Illumina recommends the use of CAT-6 cables (minimum requirement is CAT-5e). A shielded network cable that is 3 meters (9.8 feet) long is provided with the instrument for network connections. |
For the system to function properly, reserve the IP ranges 169.254/16 and IPv6 fddc:65e5:66fa::*.
Network Illustration
NOTE
The CE is not visible on the host network.
|
Connection |
Value |
Purpose |
|---|---|---|
|
Domain |
localhost:* |
All ports for localhost-to-localhost communication, which are needed for interprocess communication. |
| Port |
5555 |
Hardware controller interface |
|
9030 |
Real-Time Analysis |
|
|
8080 |
NovaSeq Operating Software |
|
|
29644 |
Universal Copy Service |
|
|
22, 80, 111, 443, 623, 2049, 5900, 8889, 9980, fddc:65e5:66fa::1/48, fddc:65e5:66fa::2/48 |
Data transfer |
|
|
29000 |
Instrument orchestrator |
Outbound connection information includes domain and IP address information for configuring access to BaseSpace Sequence Hub domains, Illumina Proactive, software updates, and uploads of run and performance data.
|
Component |
TCP |
UDP |
IP |
|---|---|---|---|
|
BMC |
22,80,443,623,5900,8889 |
623 |
169.254.0.2 |
|
NFS |
111,2049 |
111,2049 |
fddc:65e5:66fa::2/48 fddc:65e5:66fa::1/48 |
|
CE |
22,9980 |
n/a |
169.254.0.1 |
|
Hardware Controllers |
n/a |
n/a |
169.254.x.x/16 |