Select and Start Run
The instrument scans the library tube ID and searches for a matching planned run.
|
1.
|
If a planned run matching the library tube ID is found for each side being used, run selection is skipped. Select Review to proceed. |
|
2.
|
If there is no matching run for one or either side, select Run Selection, and then select one or more planned runs. |
The same planned run cannot be selected on both sides.
|
3.
|
When one or more runs are selected, select Pre-Run Checks. |
|
4.
|
Wait approximately 5 minutes for the pre-run check to complete. |
The run starts automatically after successful completion.
NOTE
To avoid overfilling the hard drive, do not copy any data to C:\ after the run starts.
Pre-Run Check Errors
Refer to Troubleshooting for more information on pre-run check errors.
|
1.
|
If pre-run checks fail due to a sensor error, such as flow cell not detected, you must exit and restart the workflow. |
|
2.
|
For other pre-run check failures, select Retry to restart the failed check or Retry All to restart all checks. |
Errors require resolution before the run can start.
|
3.
|
Select the Error icon to see error details. |
|
4.
|
If the alignment check fails, resolve the error as follows. |
|
a.
|
Select Reload, and then select OK to return to the Load screen. |
|
b.
|
Remove any items from the top of the instrument, and then select OK. The flow cell door opens. |
|
c.
|
Reload the flow cell, and then select Run Setup. |
|
d.
|
Proceed through each screen to reread each RFID and return to the Pre-Run Checks screen. |