Select and Start Run

The instrument scans the library tube ID and searches for a matching planned run.

1. If a planned run matching the library tube ID is found for each side being used, run selection is skipped. Select Review to proceed.
2. If there is no matching run for one or either side, select Run Selection, and then select one or more planned runs.

The same planned run cannot be selected on both sides.

3. When one or more runs are selected, select Pre-Run Checks.
4. Wait approximately 5 minutes for the pre-run check to complete.

The run starts automatically after successful completion.

NOTE

To avoid overfilling the hard drive, do not copy any data to C:\ after the run starts.

Pre-Run Check Errors

Refer to Troubleshooting for more information on pre-run check errors.

1. If pre-run checks fail due to a sensor error, such as flow cell not detected, you must exit and restart the workflow.
2. For other pre-run check failures, select Retry to restart the failed check or Retry All to restart all checks.

Errors require resolution before the run can start.

3. Select the Error icon to see error details.
4. If the alignment check fails, resolve the error as follows.
a. Select Reload, and then select OK to return to the Load screen.
b. Remove any items from the top of the instrument, and then select OK. The flow cell door opens.
c. Reload the flow cell, and then select Run Setup.
d. Proceed through each screen to reread each RFID and return to the Pre-Run Checks screen.
e. Redo the check.