NextSeq 1000/2000 Integration v2.1 Validation and Troubleshooting
This section explains how to validate the installation of the Illumina NextSeq 1000/2000 Integration Package v2.1.
The validation process involves:
| • | Running samples through the Library Prep Validation v2.0.3 workflow. |
| – | The workflow contains a single-step protocol that models the library prep workflow required to produce libraries tagged with index sequences. At the end of the step, these libraries are routed to NextSeq 1000/2000 Sequencing v2.0 workflow. |
| • | Running the libraries through the NextSeq 1000/2000 Sequencing v2.0 workflow. This process validates the following information: |
| – | Successful sequential step advancement of samples in the following steps: |
| 1. | Library Pooling and Dilution (NextSeq 1000/2000 Sequencing v2.0) |
| 2. | Load To Reagent Cartridge (NextSeq 1000/2000 Sequencing v2.0) |
| 3. | AUTOMATED - Sequencing Run (NextSeq 1000/2000 Sequencing v2.0) |
| 4. | [Optional] Demultiplexing (NextSeq 1000/2000 Sequencing v2.0) |
| – | Automatic validation of run setup information before sample sheet generation or planned run creation. |
| – | Automated generation of a sample sheet. The sample sheet is used to start the sequencing run on NextSeq 1000/2000 Control Software via Local run mode. |
| – | Automated creation of a planned run on Illumina Connected Analytics (ICA). The planned run created is retrieved by NextSeq 1000/2000 Control Software to start run via Hybrid or Cloud run mode. |
| – | Automated tracking of the NextSeq 1000/2000 sequencing run and parsing of run statistics from BaseSpace Sequence Hub (BSSH) into Clarity LIMS using SIS Integration Service. |
| – | Automated tracking of the secondary analysis using DRAGEN (up to BCL Convert only) and parsing of demultiplexing metrics from BaseSpace Sequence Hub (BSSH) into Clarity LIMS using SIS integration service. |
Before running the validation steps below, these steps assume that the NextSeq 1000/2000 Sequencing v2.0 Integration Package is installed and you have imported the default Clarity LIMS configuration.
For information on how the integration works, refer to NextSeq 1000/2000 Integration v2.1 Configuration and User Interaction.