AmpliSeq for Illumina Focus Panel
Protocol 1: DNA Library Prep AmpliSeq for Illumina Focus Panel
Master Step 1: Dilute RNA (AmpliSeq for Illumina)
Step Type: Standard
Input = RNA with concentration
Output = Diluted RNA (1-100 ng)
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
AmpliSeq Library PLUS for Illumina |
Illumina |
24 - 20019101; 96 - 200191102; 384 - 200191103 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Desired Concentration (ng/µl) |
Numeric |
Range = 20–50 |
|||
|
Multiline Text Fields |
Directions |
Default Read Only
|
|||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Global field - derived sample |
|||
|
Conc. Units |
Text |
Global field - derived sample |
|||||
|
Input Amount (ng) |
Numeric |
Global field - derived sample |
|||||
|
Nuclease-free water (µl) |
Numeric |
Global field - derived sample |
|||||
|
Sample Volume (µl) |
Numeric |
Global field - derived sample |
|||||
|
Total Volume (µl) |
Numeric |
Global field - derived sample |
|||||
|
Desired Concentration (ng/uL) |
Numeric |
Set by EPP |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Concentration and Set Sample Volume |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Input Amount |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Copy Desired Concentration |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
||||||
|---|---|---|---|---|---|---|---|
|
Start Next Step |
Manual |
||||||
|
Assign Next Step |
Manual |
||||||
|
Safe Stopping Point |
Optional
|
Master Step 2: Qubit (AmpliSeq for Illumina)
Step Type: Standard QC
Input = Diluted RNA (1-100 ng)
Output = Measeurment
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Criteria 1 - Source Data Field |
Text |
Default
|
|||
|
Criteria 1 - Operator |
Text |
Default
|
|||||
|
Criteria 1 - Threshold Value |
Numeric |
|
|||||
|
Criteria 2 - Source Data Field |
Text |
Default
|
|||||
|
Criteria 2 - Operator |
Text |
Preset
|
|||||
|
Criteria 2 - Threshold Value |
Numeric |
|
|||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||
|
QC Log File |
Auto |
||||||
|
QC Result File |
Auto |
||||||
|
Upload File |
Manual |
||||||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Global field - derived sample & measurement |
|||
|
Conc. Units |
Text |
Global field - derived sample & measurement |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Assign QC Flags |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step and Copy Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
Master Step 3: Dilute RNA (AmpliSeq for Illumina)
Step Type: Standard
Input = Diluted RNA with concentration
Output = Diluted RNA (1-100 ng)
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Desired Concentration (ng/µl) |
Numeric |
Range = 20–50 |
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Global field - derived sample |
|
Conc. Units |
Text |
Global field - derived sample |
||
|
Input Amount (ng) |
Numeric |
Global field - derived sample |
||
|
Nuclease-free water (µl) |
Numeric |
Global field - derived sample |
||
|
Sample Volume (µl) |
Numeric |
Global field - derived sample |
||
|
Total Volume (µl) |
Numeric |
Global field - derived sample |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Concentration and Set Sample Volume |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Input Amount |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step - Advance and Set Desired Conc |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
Master Step 4: Reverse Transcribe RNA (AmpliSeq for Illumina)
Step Type: Standard
Input = RNA with concentration
Output = Reverse Transcribed RNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
AmpliSeq cDNA Synthesis for Illumina |
Illumina |
20022654 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Preset
|
||||||
|
5X AmpliSeq cDNA Reaction Mix (µl) |
Numeric |
|||||||||
|
10X AmpliSeq RT Enzyme Mix (µl) |
Numeric |
|||||||||
|
Nuclease-free water (µl) |
Numeric |
|||||||||
|
Prep Input Volume (µl) |
Numeric |
Calculated |
||||||||
|
Prep Input Amount (ng) |
Numeric |
Required
|
||||||||
|
Total Samples |
Numeric |
Default
|
||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Global field - derived sample |
||||||
|
Conc. Units |
Text |
Global field - derived sample |
||||||||
|
Total Volume (µl) |
Numeric |
Global field - derived sample |
||||||||
|
Input Amount (ng) |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Concentration, Total Volume and Input Amount - Childhood Cancer Panel |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
Master Step 5: Amplify cDNA Targets Standard (AmpliSeq for Illumina)
Step Type: standard
Input = Reverse Transcribed RNA
Output = cDNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
AmpliSeq Library PLUS for Illumina |
Illumina |
24 - 20019101; 96 - 200191102; 384 - 200191103 |
|
AmpliSeq Focus Panel for Illumina |
Illumina |
20019164 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Default
|
|||
|
5X AmpliSeq HiFi Mix (uL) |
Numeric |
Decimal Places = 2 |
|||||
|
Nuclease-free water (µl) |
Numeric |
Decimal Places = 2 |
|||||
|
PCR Cycles |
Numeric |
||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples and Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 6: Partially Digested Amplicons (AmpliSeq for Illumina)
Step Typ = No Outputs
Input = Pooled samples
Output = Partially Digested Amplicons
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
AmpliSeq Library PLUS for Illumina |
Illumina |
24 - 20019101; 96 - 200191102; 384 - 200191103 |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Preset
|
|||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing cDNA- Focus Panel |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 2: DNA Library Prep (AmpliSeq for Focus Panel)
Master Step 1: Dilute DNA (AmpliSeq for Illumina)
Step Type: Standard
Input = DNA with concentration
Output = Diluted DNA (1-100 ng)
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
AmpliSeq Library PLUS for Illumina |
Illumina |
24 - 20019101; 96 - 200191102; 384 - 200191103 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Desired Concentration (ng/µl) |
Numeric |
Range = 20–50 |
|||
|
Multiline Text Fields |
Directions |
Multiline Text |
Default Read Only
|
||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Global field - derived sample |
|||
|
Conc. Units |
Text |
Global field - derived sample |
|||||
|
Input Amount (ng) |
Numeric |
Global field - derived sample |
|||||
|
Low TE (µl) |
Numeric |
Global field - derived sample |
|||||
|
Sample Volume (µl) |
Numeric |
Global field - derived sample |
|||||
|
Total Volume (µl) |
Numeric |
Global field - derived sample |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Concentration and Set Sample Volume |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Input Amount |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|
Setting |
Value |
||||||
|---|---|---|---|---|---|---|---|
|
Start Next Step |
Manual |
||||||
|
Assign Next Step |
Manual |
||||||
|
Safe Stopping Point |
Optional
|
Master Step 2: Qubit (AmpliSeq for Illumina)
Step Type: Standard QC
Input = Diluted DNA (1-100 ng)
Output = Measurement
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Criteria 1 - Source Data Field |
Text |
Default
|
|||
|
Criteria 1 - Operator |
Text |
Default
|
|||||
|
Criteria 1 - Threshold Value |
Numeric |
|
|||||
|
Criteria 2 - Source Data Field |
Text |
Default
|
|||||
|
Criteria 2 - Operator |
Text |
Default
|
|||||
|
Criteria 2 - Threshold Value |
Numeric |
|
|||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||
|
QC Log File |
Auto |
||||||
|
QC Result File |
Auto |
||||||
|
Upload File |
Manual |
||||||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Global field - derived sample |
|||
|
Conc. Units |
Text |
Global field - derived sample |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Assign QC Flags |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step and Copy Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
Master Step 3: Dilute DNA (AmpliSeq for Illumina)
Step Type: Standard
Input = DNA with concentration
Output = Diluted DNA (1-100 ng)
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
AmpliSeq Library PLUS for Illumina |
Illumina |
24 - 20019101; 96 - 200191102; 384 - 200191103 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Desired Concentration (ng/µl) |
Range = 20–50 |
|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Global field - derived sample |
|
Conc. Units |
Text |
Global field - derived sample |
||
|
Input Amount (ng) |
Numeric |
Global field - derived sample |
||
|
Low TE (µl) |
Numeric |
Global field - derived sample |
||
|
Sample Volume (µl) |
Numeric |
Global field - derived sample |
||
|
Total Volume (µl) |
Numeric |
Global field - derived sample |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Concentration and Set Sample Volume |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Input Amount |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 4: Amplify Targets Duplicates (AmpliSeq for Illumina)
Step Type: Standard
Input = Diluted DNA (1-100 ng)
Output = Amplified DNA
Sample Name = {InputItemName}
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
AmpliSeq Library PLUS for Illumina |
Illumina |
24 - 20019101; 96 - 200191102; 384 - 200191103 |
|
AmpliSeq Focus Panel for Illumina |
Illumina |
20019164 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Default
|
||||||
|
PCR Cycles |
Numeric |
|||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
5X AmpliSeq HiFi Mix (µl) |
Numeric |
Global field - derived sample |
||||||
|
Concentration |
Numeric |
Global field - derived sample |
||||||||
|
Conc. Units |
Text |
Global field - derived sample |
||||||||
|
Nuclease-free water (µl) |
Numeric |
|
||||||||
|
Prep Input Volume (µl) |
Numeric |
|
||||||||
|
Prep Input Amount (ng) |
Numeric |
Required
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Concentration, Total Volume and Input Amount- Focus Panel |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 6: Partially Digested Amplicons (AmpliSeq for Illumina)
Step Type: No Outputs
Input = Pooled samples
Output = Partially Digested Amplicons
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
AmpliSeq Library PLUS for Illumina |
Illumina |
24 - 20019101; 96 - 200191102; 384 - 200191103 |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Default
|
|||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 5: Transfer RNA and DNA Amplicons (AmpliSeq for Illumina)
Step Type: Standard
Input = cDNA and DNA amplicon samples
Output = cDNA and DNA samples
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
Proof of concept on API page for automatic plating.
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 6: Ligate Indexes (AmpliSeq for Illumina)
Step Type: Add Labels
Input = Partially Digested Amplicons
Output = Ligated Libraries
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
AmpliSeq Library PLUS for Illumina |
Illumina |
24 - 20019101; 96 - 200191102; 384 - 200191103 |
|
Setting |
Value |
|---|---|
|
Destination Containers |
96 well plate |
|
Setting |
Value |
|---|---|
|
Label Groups |
AmpliSeq CD Indexes Set A for Illumina - AmpliSeq CD Indexes Set A for Illumina-label-sheet.xlsx |
|
AmpliSeq CD Indexes Set B for Illumina - AmpliSeq CD Indexes Set B for Illumina-label-sheet.xlsx |
|
|
AmpliSeq CD Indexes Set C for Illumina - AmpliSeq CD Indexes Set C for Illumina-label-sheet.xlsx |
|
|
AmpliSeq CD Indexes Set D for Illumina - AmpliSeq CD Indexes Set D for Illumina-label-sheet.xlsx |
|
|
AmpliSeq UD Indexes for Illumina (24 Indexes) - AmpliSeq UD Indexes for Illumina (24 Indexes)-label-sheet.xlsx |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Preset
|
|||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||
|
Sample Table |
Table Columns |
AmpliSeq Workflow |
Text Dropdown |
Required
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Route AmpliSeq Samples - Focus Panel |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 3: Standard Workflow (AmpliSeq for Focus Panel)
Master Step 1: Clean Up Library (AmpliSeq for Illumina)
Step Type: No Outputs
Input = Ligated Libraries
Output = Cleaned Up Libraries
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Agencourt AMPure XP Beads |
Thermo |
NC9959336; NC9933872 |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
70% EtOH Prep Date |
Date |
|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 2: Amplify Library (AmpliSeq for Illumina)
Step Type: No Outputs
Input = Cleaned Up Libraries
Output = Amplified Libraries
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Agencourt AMPure XP Beads |
Thermo |
NC9959336; NC9933872 |
|
AmpliSeq Library PLUS for Illumina |
Illumina |
24 - 20019101; 96 - 200191102; 384 - 200191103 |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Group of Defaults
|
|||||||
|
Total Samples |
Numeric |
Default
|
||||||||
|
1X Lib AMP Mix (µl) |
Numeric |
|||||||||
|
10X Library Amp Primers (µl) |
Numeric |
|||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples and Calculate Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 3: Perform Second Clean Up (AmpliSeq for Illumina0)
Step Type: Standard
Input = Amplified Libraries
Output = Cleaned Up Amplified Libraries
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Agencourt AMPure XP Beads |
Thermo |
NC9959336; NC9933872 |
|
AmpliSeq Library PLUS for Illumina |
Illumina |
24 - 20019101; 96 - 200191102; 384 - 200191103 |
|
Setting |
Value |
|---|---|
|
Destination Containers |
96 well plate |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
70% EtOH Prep Date |
Date |
|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
End of Standard Protocol
Protocol 4: Equalizer Workflow (Ampliseq for Illumina Focus Panel)
Master Step 1: Clean Up Library (AmpliSeq for Illumina)
Step Type: No Outputs
Input = Ligated Libraries
Output = Cleaned Up Libraries
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Agencourt AMPure XP Beads |
Thermo |
NC9959336; NC9933872 |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
70% EtOH Prep Date |
Date |
|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 2: Amplify Library (AmpliSeq for Illumina)
Step Type: No Outputs
Input = Cleaned Up Libraries
Output = Amplified Libraries
|
Kit Name |
Supplier |
|---|---|
|
Agencourt AMPure XP Beads |
|
|
AmpliSeq Library PLUS for Illumina |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Group of Defaults
|
||||||
|
Total Samples |
Numeric |
Default
|
||||||||
|
1X Lib AMP Mix (µl) |
Numeric |
|||||||||
|
10X Library Amp Primers (µl) |
Numeric |
|||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples and Calculate Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 3: Perform Capture and Clean Up (AmpliSeq for Illumina)
Step Type: Standard
Input = Amplified Libraries
Output = Cleaned Up Libraries
|
Kit Name |
Supplier |
|---|---|
|
AmpliSeq Library Equalizer for Illumina |
Illumina |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 4: Elute Library (AmpliSeq for Illumina)
Step Type: Standard
Input = Cleaned Up Libraries
Output = Final Library
|
Kit Name |
Supplier |
|---|---|
|
AmpliSeq Library Equalizer for Illumina |
Illumina |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Default
|
|||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
End of Equalizer ProtocoL