Switch Remote Reader to Local Reader

If you have a dual reader configuration and must take the local reader offline, you can use AutoLoader 2.x to load your samples on the other reader. AutoLoader 2.x can switch the remote reader to be the new local reader.

AutoLoader 2.x software and AutoLoader 2.x Service software must be installed on both readers to switch the local and remote readers.

In a dual-reader configuration, a DB-9 switchbox typically sits on the back left side of the AutoLoader 2.x base plate, next to the left-side reader. The switchbox determines which reader is remote and which is local.

The DB-9 switchbox is only provided for flexibility. It is not recommended to switch local and remote readers regularly. If you must switch readers, make sure you follow the instructions in this section or the readers might not work properly.