Power Up AutoLoader 2.x
Power up AutoLoader 2.x as follows.
Turn on AutoLoader 2.x only after the iScan reader has been turned on.
The AutoLoader 2.x is designed for safe operation. However, any system with moving parts can potentially cause injury or damage equipment. Do not touch the robot arm of AutoLoader 2.x while it is moving. Do not put your face, body, or any lab equipment in its path.
| 1. | With the AutoLoader 2.x power turned off, move the robot arm to rest on top of the lid stack. |
Moving the arm prevents it from contacting the reader tray when AutoLoader 2.x is turned on and initialized.
| 2. | Make sure that AutoLoader 2.x has been off for at least 2 minutes. |
| 3. | Press the power switch on the back panel. |