Perform this activity annually or whenever mold or bacterial growth is detected in the robot fluidics.
|
1.
|
In the carboy, prepare a 1 L 10% bleach solution by adding 100 ml bleach to 900 ml DI H2O. |
|
2.
|
Select Robot QC Tasks | Robot Bleach Wash. |
|
3.
|
In the Basic Run Parameters pane, enter 1 for a system wash. |
|
5.
|
When prompted, swap out the system liquid carboy for the carboy containing 10% bleach solution. |
|
7.
|
When prompted, swap out the bleach solution with the system liquid carboy. |
|
9.
|
Observe lines for air bubbles. |
|
10.
|
Run additional washes (SysWash) until the lines are free of air bubbles. If bubbles persist in the lines, contact Illumina Technical Support. |
To prevent contamination, make sure that all bleach vapors have fully dissipated before starting any process involving samples.