Troubleshooting
Refer to the following solutions and workarounds if the Infinium Assay assay operates in an unexpected manner. If the issue cannot be resolved using this guide, contact Illumina Technical Support.
|
# |
Symptom |
Probable Cause |
Resolution / Comment |
|---|---|---|---|
|
1 |
Make sure to centrifuge the reagent tubes at 280 × g after thawing. |
Check pipette calibration. A gravimetric test using water is a quick and easy way to check for accurate pipette dispensing volumes. Recalibrate pipettes yearly. |
|
|
Programmable pipettor may be set incorrectly. |
Check to see if the programmable pipettors have an overasp volume that is discarded. |
||
|
Excessive evaporation after heat denaturing. |
Foil heat sealer should have been used. Ensure proper seating. |
||
|
2 |
Cap mat deformed or melted while heat-denaturing DNA samples, non-EX assays only. |
Foil heat sealer should have been used. |
Samples have been ruined. Repeat experiment. Use the foil heat sealer for all temperatures ≥ 45°C. |
|
3 |
A small amount of precipitate was present in the hybridization solution. |
A small amount of precipitate is normal, and does not affect data quality. |
Continue with experiment. |
|
4 |
Large precipitate cannot be resuspended. |
Excessive evaporation after heat denaturing. |
Foil heat sealer should have been used. Ensure proper sealing. |
|
5 |
BeadChips are still wet on the underside after 55 minutes in the vacuum desiccator. |
The BeadChips must dry for a longer period. XC4 may be old and must be replaced with fresh XC4. An old bottle of ethanol may have absorbed atmospheric water. Replace with a fresh bottle of ethanol. |
Dry BeadChips longer under vacuum desiccator. Lab temperature and humidity affect drying time. |
|
6 |
After coating the BeadChips with XC4, some uncoated areas remain. |
During the coating process, a bubble formed between the BeadChips and prevented the XC4 solution from reaching the surface. |
Briefly place the staining rack with BeadChips back into the wash dish containing XC4. Gently move BeadChips back and forth while moving up and down, breaking the surface of the solution. The back-and-forth movement is especially important when processing 16 or 24 BeadChips. |
|
7 |
During XStain, the liquid in the Flow-Through Chamber dropped below the bottom edge of the glass back plate reservoir. |
Glass back plates may not have been completely clean, causing capillary gap failure. Make sure that the correct spacer was used to assemble the Flow Through Chamber. |
Before reassembling the Flow-Through Chambers, clean the glass back plates. |
|
Attach the metal clamps to the Flow-Through Chambers as described in the Reference Guide for the assay. |
|
# |
Symptom |
Probable Cause |
Resolution / Comment |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
1 |
The iScan was unable to find all the fiducials during scanning. |
XC4 coating was not properly removed from BeadChip edges. |
Rewipe the edges of BeadChips with ProStat EtOH wipes and rescan. |
||||||||||||
|
BeadChips were not seated correctly in BeadChip carrier. |
Reseat BeadChips in BeadChip carrier. |
||||||||||||||
|
2 |
The absence of an intensity image may be due to failures in any experimental steps upstream of scanning. |
Repeat the experiment. Check the staining controls. If the controls do not generate a signal either, then the staining reagents may be compromised. |
|||||||||||||
|
3 |
The scanning process generated a low assay signal. The Hyb controls display expected intensities. |
A low assay accompanied by normal performance from the sample-independent controls signal indicates a sample-dependent failure. The error may have occurred in any experimental steps between amplification and hybridization. |
Repeat experiment. Make sure that there is a DNA pellet after precipitation. During resuspension, make sure that the DNA pellet dissolves properly. The blue color in the pellet should disappear. |
||||||||||||
|
4 |
The scanning process generated a low assay signal. The Hyb controls display low intensities. |
A low assay signal accompanied by low sample-independent controls indicates a sample-independent failure related to assay processing. The error probably occurred after hybridization. |
To eliminate the scanner as the problem, make sure that other chips that are scanned at the same time or in the same batch are yielding normal assay signals. If other chips show the same low signal, contact Illumina Technical Support to analyze the data in GenomeStudio to determine possible root causes. |
||||||||||||
|
5 |
Some spots on the BeadChip have low to 0 intensity. |
Bubbles in the reagents can cause this phenomenon by preventing reagents from reaching the BeadChip surface. |
Centrifuge all reagent tubes before using to prevent bubbles. Always perform a system flush before running the experiment. (Low to 0 intensity areas on the BeadChip may not have a negative effect on array data due to randomness and oversampling of BeadChips.) |
||||||||||||
|
6 |
The scanner returns red stripes indicating a failed scan. |
|
Move the chip to another position in the carrier and rescan. If this fails, contact Illumina Technical Support. |
|
# |
Symptom |
Probable Cause |
Resolution / Comment |
|---|---|---|---|
|
1 |
Genotyping results do not correlate with samples. |
The samples may have been mixed up. |
Check the sample sheet and lab tracking form to determine the location of DNA samples during amplification. |
|
Check to make sure that the correct sample was loaded onto the correct BeadChip or part of the BeadChip. |
|||
|
Repeat the experiment. After resuspending the DNA pellet, make sure to consolidate the samples back to the original amplification position, if necessary. |
|||
|
2 |
Pipetting may have been too vigorous. |
Pipette gently without forming bubbles. Try centrifuging the plate to 280 × g to remove bubbles. |
|
|
3 |
Heat block cooled to less than 37°C when it was set to 37°C after the 95°C incubation. |
Heat block overshot set temperature. |
Wait for heat block to reequilibrate to 37°C. Many heat blocks will overshoot the bottom set temperature when cooling from a previously set higher temperature. |
