Environmental Considerations
This snippet causes dropdowns to open on page load. This text/paragraph does not appear in the output.
The iScan is for indoor use only.
|
Element |
Specification |
|---|---|
|
Temperature |
Transportation and Storage: -10°C to 50°C (14°F to 122°F) Operating Conditions: 15°C to 30°C (59°F to 86°F) During a run, do not allow the ambient temperature to vary by more than ±2°C. |
|
Humidity |
Transportation and Storage: 15–80% relative humidity (non-condensing) Operating Conditions: 20–80% relative humidity (non-condensing) |
|
Altitude |
Locate the instrument at an altitude no higher than 2000 meters (6500 feet). |
|
Air Quality |
Operate the instrument in a Pollution Degree II environment or better. A Pollution Degree II environment normally includes only nonconductive pollutants. |
|
Ventilation |
Consult your facilities department for ventilation requirements sufficient for the level of heat output expected from the instrument. |
The following table shows the heat output for one instrument with a personal computer (PC).
|
Component |
Measured Power (Watts) |
Thermal Output (Btu/h) |
|---|---|---|
|
Instrument |
750 |
2600 |
|
A8202 PC and Monitor |
400 |
1400 |
|
Total Estimated Heat Output |
1150 |
4000 |