Verify DMAP File Integrity
If bead map (DMAP) files transfer incorrectly to the network during downloading, file integrity can become compromised. The iScan Reader can be configured to check the integrity of DMAP files at the start of each scan.
Enabling this feature increases the amount of time before the iScan Reader begins performing the scan.
| 1. | From the menu in the upper left corner of the screen, select Tools, and then select Options. |
| 2. | In the Options dialog box, select the General tab. |
| 3. | In the Processing section, select the Enable Corrupt DMAP Check checkbox, and then select OK. |