Laboratory Requirements
This snippet causes dropdowns to open on page load. This text/paragraph does not appear in the output.
Use the following specifications and guidelines to determine required lab space.
The iScanDx and instrument control computer have the following dimensions after installation.
|
Measurement |
iScanDx |
Instrument Control Computer |
AutoLoader 2.x |
|---|---|---|---|
|
Width |
52 cm (20.5 in) |
21.6 cm (8.5 in) |
85 cm (33.4 in) |
|
Height |
46 cm (18 in) |
57 cm (22.3 in) |
76 cm (29.9 in) |
|
Depth |
66 cm (26 in) without keyboard tray 85 cm (33.5 in) with keyboard tray |
55.4 cm (21.8 in) |
65 cm (25.5 in) |
|
Weight |
70 kg (154.3 lb) |
34 kg (75 lb) |
Refer to AutoLoader weight table. |
The AutoLoader 2.x has the following weight values, based on its configuration:
|
Configuration |
AutoLoader 2.x Weight Without iScanDx |
AutoLoader 2.x Weight With iScanDx |
|---|---|---|
|
Single AutoLoader 2.x (10 Carriers) |
85 kg (187.4 lb) |
155.8 kg (343.4 lb) |
|
Dual AutoLoader 2.x (20 Carriers) |
125 kg (275.6 lb) |
266.6 kg (587.8 lb) |
|
BeadChip Carrier |
0.25 kg (0.55 lb) |
0.25 kg (0.55 lb) |
A dual-scanner configuration, consisting of two iScanDx systems and an AutoLoader 2.x, fits on a typical lab bench. For more information, refer to Lab Bench Guidelines.
Dual-Scanner Configuration Dimensions with AutoLoader 2.x
The following requirements ensure access to the instrument power switch and power outlet, allow proper ventilation, and provide sufficient access for servicing the instrument.
| • | Allow for an additional 15.5 cm (6 in) behind the instrument. |
| • | Position the iScanDx so that personnel can quickly disconnect the power cord from the outlet. |
| • | Allow at least 42 cm (17 in) above the iScanDx loaded on the work bench so that the exterior panels can be removed during installation and maintenance. The height dimension listed in the following laboratory space planning table includes this required overhead space. |
|
Measurement |
Instrument |
PC, Keyboard, and Monitor |
|---|---|---|
|
Width |
112 cm (44 in) |
46 cm (18 in) |
|
Height |
94 cm (37 in) |
Lab defined |
|
Depth |
101 cm (40 in) |
40 cm (16 in) |
It is recommended that the bench top vibrations stay below the vibration criterion (VC)-E curve. Make sure that there are minimal disturbances or sources of vibration on or near where the iScanDx is installed. Use the following best practices to minimize vibrations:
| • | Do not place other equipment on the bench that can produce vibrations that exceed the VC-E curve. |
| • | Do not place objects on top of the instrument. |
| • | Keep sources of vibration away from the instrument. |
| • | Only use the keyboard and mouse to interact with the instrument during operation. |
Place the instrument on a mobile lab bench with locking casters. The lab bench surface must be level (within ± 2 degrees) and nonvibrational. The bench must support the weight of the instrument and instrument control computer.
|
Width |
Height |
Depth |
|---|---|---|
|
152.4 cm (60 in) |
76.2–91.4 cm (30–36 in) |
76.2 cm (30 in) |
If located in North America, Illumina recommends the Bench-Craft mobile bench with casters (PN HS-30-60-30 P2) or equivalent.