Liquid Handling Arm Specifications

The following table lists the precise specifications regarding LiHa repeatability and positioning accuracy at 9 mm spacing, with all eight tips operating simultaneously.

Worn parts can result in deterioration of repeatability and accuracy, which affects these specifications.

LiHa Positioning Accuracy and Repeatability Specifications

Axis

Accuracy

Repeatability

X

±0.4 mm (0.016 in)

±0.15 mm (0.006 in)

Y

±0.4 mm (0.016 in)

±0.15 mm (0.006 in)

Z

±0.4 mm (0.016 in)

±0.3 mm (0.012 in)

While the Infinium workflow does not use this capability, each tip can independently detect the surface of a conductive liquid by measuring changes in capacitance. Each channel incorporates individual liquid detection. In general, this detection is feasible for conductive liquids of the following volumes:

≥ 50 μl: Low-conductive liquid in microplates with round bottoms.
≥ 100 μl: Conductive liquid in sample tubes with a diameter of 10 mm or 13 mm (0.39 in or 0.51 in).
≥ 150 μl: Conductive liquid in sample tubes with a diameter of 16 mm (0.63 in).
≥ 5 ml: Conductive liquid in a reagent trough.