Custom Primers
Illumina primers are in positions 12, 13, and 14 of the MiSeq reagent cartridge.
|
Reservoir |
Read |
|---|---|
|
12 |
Read 1 Primer Mix |
|
13 |
Index Primer Mix |
|
14 |
Read 2 Primer Mix |
You can use custom sequencing primers for a run on the MiSeq. To do so, specify the use of custom primers in the sample sheet, and then load prepared custom primers into the empty reservoirs on the MiSeq reagent cartridge. The empty reservoirs are labeled 18, 19, and 20, and each reservoir is associated with a specific read in a sequencing run.
|
Reservoir |
Read |
|---|---|
|
18 |
Read 1 custom primer |
|
19 |
Index Read custom primer |
|
20 |
Read 2 custom primer |
You can use a combination of custom primers and standard Illumina primers provided in the reagent cartridge. For example, you can use the Illumina primer for Read 1, a custom primer for the Index Read, and the Illumina primer for Read 2.
For more information about PhiX control, refer to the MiSeq System Denature and Dilute Libraries Guide (document # 15039740). When custom primers are used for Read 1 and Read 2, the software directs the sippers to reservoirs 18 and 20; therefore, Illumina primers are not used for the sequencing run. If Illumina primers are not used, the optional Illumina PhiX control cannot be sequenced. For more information about PhiX control, refer to the MiSeq System Denature and Dilute Libraries Guide (document # 15039740).