Perform Second Wash
Always use fresh wash solution for each wash step. Reusing wash solution from the previous wash can return waste to the fluidics lines.
| 1. | Prepare fresh wash solution with Tween 20 and laboratory-grade water, as follows: |
| a. | Add 5 ml 100% Tween 20 ml to 45 ml laboratory-grade water. These volumes result in 10% Tween 20. |
| b. | Add 25 ml 10% Tween 20 ml to 475 ml laboratory-grade water. These volumes result in a 0.5% Tween 20 wash solution. |
| c. | Invert five times to mix. |
| 2. | When the first wash is complete, remove the wash tray and wash bottle, and discard the remaining wash solution. |
| 3. | Refill the wash components with fresh wash solution, as follows: |
| a. | Add 6 ml wash solution to each reservoir of the wash tray. |
| b. | Add 350 ml wash solution to the 500 ml wash bottle. |
| 4. | Load the wash tray and wash bottle, as follows: |
| a. | Slide the wash tray into the reagent chiller until it stops. Close the reagent chiller door. |
| b. | Load the wash bottle and slowly lower the sipper handle, making sure that the sippers lower into the wash bottle and waste bottle. |
| c. | Close the reagent compartment door. |
| 5. | Select Next. The second wash begins. |