Measure Expected Wash Volumes
Measuring expected wash volumes confirms that wash fluidics are performing efficiently.
1. | Before beginning a wash, empty the waste bottle. |
2. | When the wash is complete, measure the wash volume in the waste bottle. |
Wash Type |
Expected Wash Volume |
---|---|
Post Run Wash |
17.25 ml |
Post Run Wash that includes a Template Line Wash |
25.5 ml |
Standby Wash |
46 ml |
Maintenance Wash |
51.75 ml |