Perform a System Check

The System Check screen provides troubleshooting options for checking the operational status of instrument components. This feature is typically used by an Illumina Technical Support representative during a Live Help session. Use of this feature is not required during normal operation or for instrument maintenance.

Some system checks can be performed before contacting Illumina Technical Support, such as the Volume Test. A volume test checks the health of the fluidics system by estimating the flow volume as bubbles pass by the sensors. For more information, refer to Perform a Volume Test.

The Tip/Tilt and Full Optics tests require a special flow cell and should only be performed by an Illumina technician.

1. From the Main Menu, select System Check.
2. Do one of the following:
Select the individual tests that you want to perform.
Choose Select All to perform all tests.
3. Select Next.

When complete, the test results appear on the screen.

4. [Optional] Select Show Details to review a summary of the results on the software interface.
5. [Optional] Select Export Results to export the results in a *.csv file format to a USB drive.
6. Select Done.