Uninterruptible Power Supply
Illumina recommends the use of a user-supplied uninterruptible power supply (UPS). Illumina is not responsible for runs affected by interrupted power regardless of whether the instrument is on a UPS. Standard generator-backed power is often not uninterruptible and a brief power outage occurs before power resumes, which interrupts a sequencing run.
Specification |
Japan APC Smart-UPS Part #SUA1500JB |
North America APC Back-UPS Pro Part # BR1500MS2 |
International APC Back-UPS Pro Part # BR1500MS2 |
---|---|---|---|
Maximum Output Power |
980 W / 1500 VA |
900 W / 1500 VA |
865 W / 1500 VA |
Input Voltage (nominal) |
100 VAC |
120 VAC |
230 VAC |
Input Connection |
NEMA 5-15P |
NEMA 5-15P |
IEC-320 C14 |
Typical Run Time (50% load) |
23.9 minutes |
14.5 minutes |
15.8 minutes |
Typical Run Time (100% load) |
6.7 minutes |
4.1 minutes |
5.5 minutes |
To obtain an equivalent UPS that complies with local standards for facilities outside the referenced regions, consult a third-party supplier such as Interpower Corporation.