Vibration Guidelines
Keep the vibration level of the lab floor at, or below, the VC-A standard of 50 μm/s for ⅓ octave band frequencies of 8–80 Hz, or lower. This level is typical for labs. Do not exceed the ISO Operating Room (baseline) standard of 100 μm/s for ⅓ octave band frequencies of 8–80 Hz.
During sequencing runs, use the following best practices to minimize vibrations and ensure optimal performance:
| • | Place the instrument on a sturdy lab bench. |
| • | Do not place keyboards, used consumables, or other objects on top of the instrument. |
| • | Do not install the instrument near sources of vibration that exceed the ISO Operating Room standard. For example: |
| – | Motors, pumps, shake testers, drop testers, and heavy air flows in the lab. |
| – | Floors directly above or below HVAC fans, and controllers, and helipads. |
| – | Construction or repair work on the same floor as the instrument. |
| – | Areas with high foot traffic. |
| • | Keep sources of vibration such as dropped items and movement of heavy equipment at least 100 cm (39.4 in) from the instrument. |
| • | Use only the touch screen, keyboard, and mouse to interact with the instrument. Do not directly impact the instrument surfaces during operation. |
