Vibration Guidelines

Keep the vibration level of the lab floor at, or below, the VC-A standard of 50 μm/s for ⅓ octave band frequencies of 8–80 Hz, or lower. This level is typical for labs. Do not exceed the ISO Operating Room (baseline) standard of 100 μm/s for ⅓ octave band frequencies of 8–80 Hz.

During sequencing runs, use the following best practices to minimize vibrations and ensure optimal performance:

Place the instrument on a sturdy lab bench.
Do not place keyboards, used consumables, or other objects on top of the instrument.
Do not install the instrument near sources of vibration that exceed the ISO Operating Room standard. For example:
Motors, pumps, shake testers, drop testers, and heavy air flows in the lab.
Floors directly above or below HVAC fans, and controllers, and helipads.
Construction or repair work on the same floor as the instrument.
Areas with high foot traffic.
Keep sources of vibration such as dropped items and movement of heavy equipment at least 100 cm (39.4 in) from the instrument.
Use only the touch screen, keyboard, and mouse to interact with the instrument. Do not directly impact the instrument surfaces during operation.