Custom Kits
You can add custom or third-party index adapter and library prep kits to the MiSeq i100 Plus Control Software. The kits are available within the on-instrument Run Planning tool during run setup.
When adding a library prep kit, you must specify one or more compatible index adapter kits. If you need to add a custom index adapter kit, add it before adding the library prep kit.
Add a Custom Index Adapter Kit
|
1.
|
Select the menu icon in the upper-left corner. |
|
2.
|
Select Settings, and then select Custom Kits. |
|
3.
|
Select Download Template to download the Index Adapter Kit template.tsv file |
|
4.
|
Open the template.tsv file using Microsoft Excel, Libre Office, or other similar spreadsheet editing software. |
|
5.
|
Follow the instructions in the template.tsv file to add the following index adapter kit information: |
|
a.
|
[IndexKit]—Overview information for the index adapter kit, including name, version, description, and index strategy. |
|
b.
|
[Resources]—Allows for providing adapter sequences for Read 1 and Read 2. Based on the values in this section, the imported file sets the index kit type as one of the following options: |
|
•
|
Fixed layout single plate. |
|
•
|
Fixed plate layout multi plate. |
|
c.
|
[Indices]— A list of indexes, including name, index sequence, and whether the index is for Index 1 or Index 2. |
Index names can include only alphanumeric characters.
|
6.
|
Remove the template instructions included in the angle brackets (< >), and then save the TSV file. |
|
7.
|
In the MiSeq i100 Plus Control Software user interface, select the drop-down menu in the upper left corner, and then select Custom Kits. |
|
8.
|
Select Import index adapter kit, then navigate to the custom index adapter kit *.tsv and select Open. |
|
9.
|
After successfully importing the custom index adapter kit, select the kit name to review and edit information. |
Add a Custom Library Prep Kit
|
1.
|
Select the menu icon in the upper-left corner. |
|
2.
|
Select Settings, and then select Custom Kits. |
|
3.
|
Select Add Library Prep Kit and enter the following information: |
|
•
|
[Optional] Description. |
|
•
|
[Optional] Organization. The company or institution that owns the custom library prep kit. The organization cannot be Illumina. |
|
•
|
From the drop-down list, select at least one compatible index adapter kit. |
|
5.
|
After successfully adding the library prep kit, select the kit name to review and edit information. |