First Time Setup

The MiSeq i100 Plus Control Software guides you though the first time setup. The following sections summarize the configuration settings to be configured during initial setup.

Do not disturb the instrument if the spinning wait cursor is displayed. Disruption of the process can result in an unrecoverable critical system error.

Administrator Account

You can only create one administrator account during the first time setup. After it is set up, you can create additional administrator accounts. For more information, refer to Add Users.

User name
Password

Instrument Nickname

[Optional] Instrument nickname

If you enter an instrument nickname, it displays at the bottom of the screen in the MiSeq i100 Plus Control Software.

Network Connection

Configuring the network connection during the first time setup procedure is optional, but recommended. If you do not configure the network, you must configure a USB or external storage. You cannot use Illumina Proactive, BaseSpace Sequence Hub, or any other cloud services until the network has been set up.

Illumina Proactive

Illumina Proactive is selected by default.

Send instrument performance data to Illumina. No sequencing data are sent.

System Checks

After the required configurations are in place, system checks are initiated to ensure the proper functioning of all the MiSeq i100 Plus components. The system checks include testing the flow cell door, internal cooling fan, and reagent loading mechanisms. Do not disturb the instrument as it undergoes the system checks. The system checks use the reusable wet and dry test cartridges, which are included with the MiSeq i100 Plus.

Load the reusable test cartridges as follows.

1. Select Next to extend the dry tray.
2. Load the dry test cartridge after the dry tray extends.
3. Select Next to retract the dry tray and extend the wet tray.
4. Load the wet test cartridge after the wet tray extends.
5. Press Next to retract the wet tray and start the system checks.

Do not adjust the trays manually. Doing so can lead to an unrecoverable critical system error.

If the system checks identify any failures, the system checks continue until all the components have been checked. A comprehensive list of the failed components is recorded in the log files. Contact Illumina Technical Support to share the log files and address any issues through troubleshooting.

After the system checks have completed, unload the reusable wet test cartridge and reusable dry test cartridge by selecting Eject Consumables from the Start screen. Store the cartridges at the ambient temperature for future use.