Uninterruptible Power Supply

Illumina recommends using a user-supplied uninterruptible power supply (UPS).

The following table shows example recommended UPS models for the MiSeq i100 Plus System.

Region

North America

Japan

International

Specifications

APC Smart UPS

750 VA LCD 120 V

Part # SMT750C

APC Smart UPS

750 VA LCD 100 V

Part # SMT750J

APC Smart UPS

750 VA LCD 230 V

Part # SMT750IC

Maximum Output Capacity

500 W / 750 VA

500 W / 750 VA

500 W / 750 VA

Input Voltage (nominal)

120 VAC

100 VAC

230 VAC

Input Frequency

50/60 Hz

50/60 Hz

50/60 Hz

Dimensions (H x W x D)

6.34 in (16.1cm) x 5.43 in (13.8cm) x 14.53 in (36.9cm)

16.7cm x 14cm x 35.9cm

16.1cm x 13.8cm x 36.9cm

Weight

27.56 lb(US) (12.5 kg)

13 kg

11.8 kg

Typical Run Time (at 300 Watts)

12 min 2 sec

12 min 2 sec

12 min 2 sec