Start a Manual Run (Generate BCL Files)

Use the following instructions to start a sequencing run that only generates BCL files. Sample sheet is optional.

1. Select Start.
2. If you are not signed in, follow the instructions provided in Sign In and Sign Out.
3. Select Generate BCL files.
4. Enter a run name.

The run name can only contain alphanumeric characters, spaces, dashes, and underscores.

5. Select Single or Paired end for the read type.
6. Enter the number of cycles performed in each read:

The total number of read cycles and index cycles cannot exceed the number of cycles specified by the reagent kit.

Read 1—Enter the number of cycles for Read 1.
Index 1—Enter the length of the index read for Index 1. For a PhiX-only run, enter 0 in both index fields.
Index 2—Enter the length of the index read for Index 2.
Read 2—Enter the number of cycles for Read 2. This value is typically the same as the Read 1 value.
7. [Optional] Select your sample sheet.
8. Select Review, and then review your run. Configure the following optional run settings as needed:
If Read First sequencing is required, deselect the Sequence Indexes First checkbox.
If using custom primers, select the appropriate custom primers checkboxes.
If the instrument is connected to the cloud, and you are logged in with your BaseSpace Sequence Hub account, select a cloud run setting.
To use a different output folder from the default, modify the output folder. You can change the default output folder in system settings.
Select a custom recipe file.
9. When finished, refer to Prepare Dry Cartridge.