Uninterruptible Power Supply
Illumina recommends using a user-supplied uninterruptible power supply (UPS).
The following table shows example recommended UPS models for the MiSeq i100 Series.
|
Region |
North America |
Japan |
International |
|---|---|---|---|
|
Specifications |
APC Smart UPS 750 VA LCD 120 V Part # SMT750C |
APC Smart UPS 750 VA LCD 100 V Part # SMT750J |
APC Smart UPS 750 VA LCD 230 V Part # SMT750IC |
|
Maximum Output Capacity |
500 W / 750 VA |
500 W / 750 VA |
500 W / 750 VA |
|
Input Voltage (nominal) |
120 VAC |
100 VAC |
230 VAC |
|
Input Frequency |
50/60 Hz |
50/60 Hz |
50/60 Hz |
|
Dimensions (H x W x D) |
6.34 in (16.1 cm) x 5.43 in (13.8 cm) x 14.53 in (36.9 cm) |
16.7 cm x 14 cm x 35.9 cm |
16.1 cm x 13.8 cm x 36.9 cm |
|
Weight |
27.56 lb (12.5 kg) |
13 kg |
11.8 kg |
|
Typical Run Time (at 300 Watts) |
12 min 2 sec |
12 min 2 sec |
12 min 2 sec |