Power Cycle the Instrument
Power cycling the instrument safely shuts down and restarts the system to restore a lost connection, align a specification, or resolve an initialization failure. Software messages indicate when to power cycle to resolve an error or warning.
| 1. | From the control software menu, select Shut Down Instrument. |
| 2. | If the system does not shut down, hold the power button on the right side of the instrument until the lights fade. |
| 3. | When the power button pulses, press the power off (Ο) side of the toggle switch on the rear panel. |
The power button might continue to pulse after the power is turned off.
Toggle Switch Location
| 4. | Wait 30 seconds. |
| 5. | Press the power on (I) side of the toggle switch. |
| 6. | When the power button pulses, wait 30 seconds, and then press it. |
Power Button Location
| 7. | Wait about 5 minutes for the operating system to load. When the operating system is loaded, log on to the system. |
The control software is launched and initializes the system. Wait about 5 minutes for system initialization. The Home screen appears when initialization is complete.