For most custom library types, you can pool the read primers together in one custom primer well and the index primers can be pooled together in the other custom primer well. For optimal base calling performance, do not combine a custom read primer with a custom index primer.
In the following steps, the volume of custom primer used depends on the custom primer stock concentration.
|
1.
|
If frozen, thaw each custom primer to be used. |
|
2.
|
If using custom or third-party libraries only, prepare custom read primers and/or custom index primers as follows. |
|
•
|
Use HT1 to dilute the custom read primer mix to yield a total volume of 600 µl with each custom read primer at 0.3 µM final concentration. |
|
•
|
Use HT1 to dilute the custom index primer mix to yield a total volume of 600 µl with each custom index primer at 0.6 µM final concentration. |
|
3.
|
If using custom or third-party libraries alongside PhiX or Illumina libraries, prepare custom read primers or custom index primers as follows. |
|
•
|
Add each custom read primer mix to 600 μl VP21 or HP21 for a 0.3 μM final concentration. |
|
•
|
Add each custom index primer mix to 600 μl VP14 or BP14 for a 0.6 μM final concentration. |