Requeue a Run
If an error displays for Status of Secondary Analysis in Process Management, you can requeue the run to perform on-instrument DRAGEN analysis again on the generated CBCL files. The original run folder must still be present on the instrument for requeue functionality. Using this requeue functionality does not requeue runs in BaseSpace Sequence Hub. To requeue in BaseSpace Sequence Hub, refer to the BaseSpace Sequence Hub support page.
|
1.
|
Update your sample sheet v2, and then save the sample sheet to a portable or mounted network drive.
|
|
2.
|
If you saved the sample sheet to a portable drive, plug the drive into a USB 3.0 port, located on both the side and back of the instrument. Gently move the instrument as needed to access the back. |
|
3.
|
Select the control software menu, and then select Process Management. |
|
4.
|
Make sure that there are no sequencing runs or on-instrument secondary analyses in progress. |
|
5.
|
Select Requeue next to the completed run to requeue. |
|
6.
|
Select Choose to navigate to the updated sample sheet, and then select Open. |