Return Consumables to Storage
Use the following instructions to store a thawed cartridge and flow cell if there is an instrument error during the instrument pre-run check before the fluidics check.
|
1.
|
Separate the flow cell from the cartridge. |
|
2.
|
Remove and discard diluted library from the reservoir (up to ~18 µl). |
Prepare a fresh dilution of the same library for the next run to avoid sample cross-contamination with residual library in the reservoir.
|
3.
|
Position the cartridge in 2°C to 8°C storage so that the label faces up and air can circulate on all sides. |
|
4.
|
Store consumables according to the following guidance. |
|
•
|
If cartridge was thawed in the refrigerator for 16 hours overnight, do not exceed 56 hours for all kit configurations. |
|
•
|
Standard SBS consumables: |
|
•
|
Do not exceed 72 hours. |
|
•
|
If 100-cycle, 200-cycle, or 300-cycle cartridge was thawed in the refrigerator for 12 hours overnight, do not exceed 60 hours. |
|
•
|
If 600-cycle cartridge was thawed in the refrigerator for 16 hours overnight, do not exceed 56 hours. |
|
5.
|
Return the flow cell to the original silver foil packaging with the desiccant. |
|
6.
|
Tape the foil packaging shut and place in 2°C to 8°C storage. |