Electrical Requirements
|
Type |
Specification |
|---|---|
|
Line Voltage |
100–240 Volts AC at 50/60 Hz |
|
Power Supply Rating |
750 Watts, maximum |
|
Type |
Specification |
|---|---|
|
Line Voltage |
24 Volts DC, 23A |
|
Power Supply Rating |
552 Watts, maximum |
Your facility must be wired with the following equipment:
| • | For 100–120 Volts AC—A 15 amp grounded, dedicated line with proper voltage and electrical ground is required. North America and Japan—Receptacle: NEMA 5-15. |
| • | For 220–240 Volts AC—A 10 amp grounded line with proper voltage and electrical ground is required. If the voltage fluctuates more than 10%, a power line regulator is required. |
|
|
The instrument has a connection to protective earth through the enclosure. The safety ground on the power cord returns protective earth to a safe reference. The protective earth connection on the power cord must be in good working condition when using this device. |
The instrument is equipped with an international standard IEC 60320 C14 receptacle and is shipped with a region-specific power cord.
Hazardous voltages are removed from the instrument only when the power cord is disconnected from the AC power source.
To obtain equivalent receptacles or power cords that comply with local standards, consult a third-party supplier such as Interpower Corporation.
Never use an extension cord to connect the instrument to a power supply.
The instrument contains no user-replaceable fuses.
A user-supplied uninterruptible power supply (UPS) is highly recommended. Illumina is not responsible for runs affected by interrupted power regardless of whether the instrument is connected to a UPS. Standard generator-backed power is often not uninterruptible and a brief power outage is typical before power resumes.
The following table lists region-specific recommendations.
|
Specification |
APC Smart UPS Part # SMT1500J (Japan) |
APC Smart UPS Part # SMT1500C (North America) |
APC Smart UPS Part # SMT1500IC (International) |
|---|---|---|---|
|
Maximum output capacity |
980 W / 1200 VA |
1000 W / 1440 VA |
1000 W / 1500 VA |
|
Input voltage (nominal) |
100 VAC |
120 VAC |
230 VAC |
|
Input frequency |
50/60 Hz |
50/60 Hz |
50/60 Hz |
|
Input connection |
NEMA 5-15P |
NEMA 5-15P |
IEC-320 C14 Schuko CEE7/EU1-16P British BS1363A |
|
Dimensions (H × W × D) |
22.5 cm × 17.2 cm × 43.9 cm (8.9 in × 6.8 in × 17.3 in) |
21.9 cm × 17.1 cm × 43.9 cm (8.6 in × 6.7 in × 17.3 in) |
21.9 cm × 17.1 cm × 43.9 cm (8.6 in × 6.7 in × 17.3 in) |
|
Weight |
26 kg (57.3 lb) |
24.6 kg (54.2 lb) |
24.1 kg (9 lb) |
|
Typical run time (at 500 Watt) |
23 minutes |
23 minutes |
23 minutes |
To obtain an equivalent UPS that complies with local standards for facilities outside the referenced regions, consult a third-party supplier such as Interpower Corporation (www.interpower.com).