Laboratory Requirements
Use the following specifications and requirements to set up the lab space.

Measurement |
Instrument Dimensions |
---|---|
Height |
60 cm (23.6 in) |
Width |
55 cm (21.7 in) |
Depth |
65 cm (25.6 in) |
Weight |
141 kg (310.9 lb) |

Position the instrument to allow proper ventilation, access to power outlet, and access for servicing the instrument.
• | Position the instrument so that personnel can quickly disconnect the power cord from the outlet. |
• | Multiple instruments placed back-to-back must be offset at least 61 cm (24 in) on each side. |
• | Make sure that hot exhaust air does not blow into the air intake of the instrument. |
• | Make sure that the instrument is accessible from all sides and clear from obstructions, so air can circulate and the instrument can be accessed and serviced. |
• | Make sure that there is enough space in front of the instrument for the keyboard. |
• | Make sure that any shelves above the instrument are ≤ 30.5 cm (12 in) deep. |
Access |
Minimum Clearance |
---|---|
Sides |
Allow at least 50.8 cm (20 in) on each side of the instrument. |
Rear |
Allow at least 11.4 cm (4.5 in) behind the instrument. |
Top |
Allow at least 61 cm (24 in) above the instrument. |
Moving the instrument improperly can affect the optical alignment and compromise data integrity. If you have to relocate the instrument, contact your Illumina representative.

The instrument includes precision optical elements. Place the instrument on a sturdy lab bench away from sources of vibration. Isolate the instrument on a moveable bench. The measurements do not include the space required for cable management.
Width |
Height |
Depth |
Casters |
---|---|---|---|
122 cm (48 in) |
91.4 cm (36 in) |
76.2 cm (30 in) |
Optional |
For North American customers, Illumina recommends the following lab bench: Bench-Tek Solutions, part # BT40CR-3048BS-PS.

Keep the vibration level of the lab floor at or below the VC-A standard of 50 μm/s for ⅓ octave band frequencies of 8–80 Hz, or lower. This level is typical for labs. Do not exceed the ISO Operating Room (baseline) standard of 100 μm/s for ⅓ octave band frequencies of 8–80 Hz.
During sequencing runs, use the following guidelines to minimize vibrations and ensure optimal performance:
• | Place the instrument on a sturdy lab bench. |
• | Do not place keyboards, used consumables, or other objects on top of the instrument. |
• | Install the instrument away from sources of vibration that exceed the ISO Operating Room standard. For example: |
– | Motors, pumps, shake testers, drop testers, and heavy air flows in the lab. |
– | Floors directly above or below HVAC fans, and controllers, and helipads. |
– | Construction or repair work on the same floor as the instrument. |
– | Areas with high foot traffic. |
• | Keep sources of vibration such as dropped items and movement of heavy equipment at least 100 cm (39.4 in) from the instrument. |
• | Use only the touch screen, keyboard, and mouse to interact with the instrument. Do not directly impact the instrument surface during operation. |