Operating System Configurations
Illumina instruments are tested and verified to operate within specifications before shipping. After installation, changes to settings can create performance or security risks.
The following configuration recommendations mitigate performance and security risks for the operating system:
| • | Configure a password that is at least 10 characters, and use local ID policies for additional guidance. Keep a record of the password. |
| – | Illumina does not keep customer login credentials. You can reset a password from the system root account or by booting into single-user mode. |
| – | Alternatively, an Illumina representative can restore the factory default configuration. This process removes all data from the system and can extend the time required for repair or maintenance. |
| • | Maintain existing privileges for preconfigured users. Make preconfigured users unavailable as needed. |
| • | The system assigns random internal IP addresses, which it uses to communicate with the hardware components. If these IP addresses are amended, or if the method of assigning the IP addresses is changed, then hardware errors can occur (including total loss of functionality). |
| • | The control computer is designed to operate Illumina sequencing systems. Web browsing, checking email, reviewing documents, and other nonsequencing activity creates quality and security problems. |