Checks for Array Scans
|
System Checks |
Recommended Action |
|---|---|
|
Doors Closed |
Make sure that the compartment doors are closed. |
|
Consumables Loaded |
Consumable sensors do not register. Make sure that each consumable is properly loaded. On the run setup screens, select Back to return to the loading step, and repeat run setup. |
|
Required Software |
Critical components of the software are missing. Perform a manual software update to restore all software components. |
|
Verify Input Files |
Make sure that the path to the cluster file and manifest file is correct and the files are present. |
|
Instrument Disk Space |
The instrument hard drive does not have sufficient disk space to perform a run. It is possible that data from a previous run did not transfer. Clear run data from the instrument hard drive. |
|
Network Connection |
The network connection has been interrupted. Check network status and the physical network connection. |
|
Network Disk Space |
Either the BaseSpace account is full or the network server is full. |
|
Imaging System |
Recommended Action |
|---|---|
|
Imaging Limits |
Contact Illumina Technical Support. |
|
Z Steps-and-Settle |
Contact Illumina Technical Support. |
|
Bit Error Rate |
Contact Illumina Technical Support. |
|
Auto-Center |
Unload the BeadChip adapter. Make sure that the BeadChip is seated in the adapter, and then reload the adapter. |
