Enter Run and Analysis Parameters in NCS (Manual Run Mode)

1. Enter a run name of your preference.
2. [Optional] Enter a library ID of your preference.
3. From the Recipe drop-down list, select a recipe. Only compatible recipes are listed.
4. Select a read type, either Single-Read or Paired-End.
5. Enter the number of cycles for each read in the sequencing run.
Read 1—Enter a value up to 151 cycles.
Read 2—Enter a value up to 151 cycles. This value is typically the same number of cycles as Read 1.
Index 1—Enter the number of cycles required for the Index 1 (i7) primer.
Index 2—Enter the number of cycles required for the Index 2 (i5) primer.

The control software confirms your entries using the following criteria:

Total cycles do not exceed the maximum cycles allowed
Cycles for Read 1 are greater than the 5 cycles used for template generation
Index Read cycles do not exceed Read 1 and Read 2 cycles
6. [Optional] If you are using custom primers, select the checkbox for the primers used. For more information, see NextSeq Custom Primers Guide (document # 15057456).
Read 1—Custom primer for Read 1.
Read 2—Custom primer for Read 2.
Index 1—Custom primer for Index 1.
Index 2—Custom primer for Index 2.
7. Set the output folder location for the current run. Select Browse to navigate to a network location.

For information on output folder requirements, see Set Output Folder Location.

8. Select Browse to navigate to a sample sheet.

Systems configured for manual mode with Run Monitoring and Storage in BaseSpace Sequencing Hub require a sample sheet.

9. Select Purge consumables for this run.

The setting purges consumables automatically after the current run.

10. Select Next.
11. [Optional] Select the Edit icon to change run parameters.
12. Select Next.