Prepare for a Manual Post-Run Wash
|
|
1 ml, diluted to 0.12%
Loaded onto the reagent wash cartridge (position #28)
|
|
|
Used to make 125 ml 0.05% Tween 20 wash solution
Loaded onto the buffer wash cartridge (center reservoir)
|
Always use a fresh dilution of NaOCl prepared within the last 24 hours. If you make a volume larger than 1 ml, store the remaining dilution at 2°C to 8°C for use within the next 24 hours. Otherwise, discard the remaining dilution of NaOCl.
|
1.
|
Combine the following volumes in a microcentrifuge tube to result in 1 ml 0.12% NaOCl: |
|
•
|
Laboratory-grade water (976 µl) |
|
2.
|
Invert the tube to mix. |
|
3.
|
Add 1 ml 0.12% NaOCl to the reagent wash cartridge. The correct reservoir is equivalent to position #28 on the prefilled cartridge. |
Load NaOCl
|
4.
|
Combine the following volumes to result in a 0.05% Tween 20 wash solution: |
Original style buffer wash cartridge
|
•
|
Laboratory-grade water (125 ml) |
|
•
|
Add 125 ml wash solution to the center reservoir of the buffer wash cartridge. |
New style buffer wash cartridge
|
•
|
Laboratory-grade water (150 ml) |
|
•
|
Add 150 ml wash solution to the center reservoir of the buffer wash cartridge. |
|
5.
|
Select Perform Wash, and then select Manual Post-Run Wash. |