Perform a Manual Wash
Manual washes are initiated from the Home screen. Wash options include the Quick Wash and the Manual Post-Run Wash.
|
Wash Types |
Description |
||||||
|---|---|---|---|---|---|---|---|
|
Quick Wash Duration: 20 minutes |
Flushes the system with a user-supplied wash solution of laboratory-grade water and Tween 20 (buffer wash cartridge).
|
||||||
|
Manual Post-Run Wash Duration: 90 minutes |
Flushes the system with a user-supplied wash solution of laboratory-grade water and Tween 20 (buffer wash cartridge) and 0.12% sodium hypochlorite (reagent wash cartridge). Required if the automatic post-run wash was not performed. |
A manual wash requires the reagent wash cartridge and buffer wash cartridge provided with the instrument, and a used flow cell. A used flow cell can be used up to 20 times for instrument washes.
Original Style Reagent Wash Cartridge and Buffer Wash Cartridge
New Style Reagent Wash Cartridge and Buffer Wash Cartridge