Electrical Requirements
Do not remove the outer panels from the instrument. There are no user-serviceable components inside. Operating the instrument with any of the panels removed creates potential exposure to line voltage and DC voltages.
|
Type |
Specification |
|---|---|
|
Line Voltage |
200–240 VAC at 50/60 Hz |
|
Peak Power Consumption |
2700 Watts |
Your facility must be wired with a minimum 15 amp grounded line with proper voltage. Requirements can vary depending on your region. Refer to Power Cords for country-specific power requirements. An electrical ground is required. If the voltage fluctuates more than 10%, a power line regulator is required.
The instrument must be connected to a dedicated circuit that must not be shared with any other equipment.
The instrument comes with an international standard IEC 60320 C20 receptacle, and is shipped with a region-specific power cord. To obtain equivalent receptacles or power cords that comply with local standards, consult a third-party supplier such as Interpower Corporation (www.interpower.com). All power cords are 2.5 m (8 ft) in length.
Hazardous voltages are removed from the instrument only when the power cord is disconnected from the AC power source.
Never use an extension cord to connect the instrument to a power supply.
|
Region |
Shipped Power Cord |
Electrical Supply |
Socket |
|---|---|---|---|
|
Australia |
AS 3112 SAA Male to C19, 15 amps
|
Line voltage: 230 VAC Minimum amp: 15 amps |
15 Amp Type I |
|
Brazil |
NBR14136 Plug to C19, 16 amps |
Line voltage: 220 VAC Minimum amp: 16 amps |
NBR 14136 Type N |
|
China |
GB2099 to C19, 16 amps
|
Line voltage: 220 VAC Minimum amp: 16 amps |
GB 1002, GB 2099, Type I |
|
European Union¹ |
Schuko CEE 7 (EU1-16p) to C19, 16 amps
|
Line voltage: 220–240 VAC Minimum amp: 16 amps |
Schuko CEE 7/3 |
|
India |
IS1293 to C19, 16 amps
|
Line voltage: 230 VAC Minimum amp: 16 amps |
BS546A Type M |
|
Israel |
IEC 60320 C19, 16 amps
|
Line voltage: 230 VAC Minimum amp: 16 amps |
SI 32 16 Amp Type H |
|
Japan |
NEMA L6-30P, 30 amps
|
Line voltage: 200 VAC Minimum amp: 30 amps |
NEMA L6-30R |
|
New Zealand |
AS 3112 SAA Male to C19, 15 amps
|
Line voltage: 230 VAC Minimum amp: 15 amps |
Dedicated 15 Amp Type I |
|
North America |
NEMA L6-20P to C19, 20 amps
|
Line voltage: < 220 VAC Minimum amp: 20 amps
Line voltage: ≥ 220 VAC Minimum amp: 16 amps |
NEMA L6-20R |
|
Singapore |
IEC60309 316P6 to C19, 16 amps
|
Line voltage: 230–250 VAC Minimum amp: 16 amps |
IEC60309 316C6 |
|
South Africa |
SANS 164-1 to C19, 16 amps
|
Line voltage: 230 VAC Minimum amp: 16 amps |
BS546A Type M |
|
Switzerland |
SEV 1011 Type 23 Plug J, 16 amps
|
Line voltage: 230 VAC Minimum amp: 16 amps |
SEV 1011 Type 23 J socket |
|
United Kingdom |
IEC60309 316P6 to C19, 16 amps
|
Line voltage: Minimum amp: 16 amps |
IEC60309 316C6 |
¹ Excepting Switzerland and the United Kingdom.
Alternatively, all regions can use IEC 60309.