Network Connections
The NovaSeq X and NovaSeq X Plus Sequencing Systems require a dedicated network connection. The systems are not intended to store run data.
An external internet connection is optional. The following operations require an external internet connection:
| • | Uploading instrument performance data to Illumina |
| • | Remote assistance from Illumina Technical Support |
Illumina does not install or provide technical support for network connections. For guidelines on how to install and configure a network connection for Illumina systems, refer to Illumina Instrument Control Computer Security and Networking.
The following additional guidelines are applicable to the NovaSeq X and NovaSeq X Plus Sequencing Systems:
| • | To connect with local networks, the instrument includes two RJ-45 copper interfaces for use with the cables provided. Direct fiber connections to the instrument are not possible. |
| • | A 10 gigabit (Gb) intranet connection to network storage is required. Lower connection speeds result in reduced instrument availability, increased data transfer times, and might also impact sequencing run performance. |
| • | Switches must be non-blocking with full duplex ports and link speed auto-negotiation turned on. |
Use the following recommended network bandwidth per instrument for connections, based on 85–90% network efficiency.
| • | 800 megabits per second (Mb/s) (primary analysis only) or ~3.5 gigabits per second (Gb/s) (primary and secondary analysis) sustained network bandwidth for storing data to local network storage. |
| • | 800 Mb/s network bandwidth for uploading primary analysis data to the cloud. |
| • | 15 Mb/s network bandwidth for run monitoring or Illumina Proactive Support only. |
Primary analysis files include BCL files and ancillary files generated by RTA4. Secondary analysis files include on-instrument DRAGEN output files.
|
Connection |
Value |
Purpose |
|---|---|---|
|
Host |
https://127.0.0.1 |
GDS via the Kubernetes Client library |
|
Port |
6443 |
GDS via the Kubernetes Client library |
|
Connection |
Value |
Purpose |
|---|---|---|
|
Port |
53 |
Domain name resolution with customer DNS servers |
|
Port |
443 |
Off-instrument control software UI or UCS |
|
URL |
lus.edicogenome.com |
DRAGEN Licensing Server |
|
Connection |
Value |
Purpose |
|---|---|---|
|
Port |
80 |
Off-instrument control software (certificate) |
|
Port |
443 |
Off-instrument control software (UI) |