Load Lyo Insert and Library Tube Strip
Before sequencing, load the lyo insert and library tube strip into the reagent cartridge as follows.
|
1.
|
Record the library tube ID on the library tube strip. The library tube ID is used when planning a sequencing run. |
|
2.
|
Uncap the library tube strip. Do not pierce the library tube strip foils. |
|
3.
|
Dispense the appropriate amount of denatured library or denatured library with PhiX into each sample tube. |
|
4.
|
Add the appropriate amount of pre-load buffer to any unused sample tubes. |
|
5.
|
After dispensing libraries, cap the library tube strip. |
Make sure that no air gaps are present at the bottom of the tubes.
|
6.
|
Centrifuge at 280 × g for 1 minute using the library tube strip adapter. |
Make sure to counterbalance the loaded library tube strip adapter with another adapter that is loaded with a library tube strip filled to 275 or 165 µl per tube.
Centrifuging at a higher speed than Illumina recommends might cause the library tube strip RFID to detach.
|
7.
|
If libraries are not collected in the bottom of the tube, repeat centrifuge. |
|
8.
|
Insert the library tube strip into the reagent cartridge and press down. |
An audible click indicates that the library tube strip is in place.
|
9.
|
Insert the lyo insert into the reagent cartridge and press down. |
An audible click indicates that the lyo insert is in place.