Prepare and Add Custom Primers
Custom primers are prepared using NovaSeq X Series Custom Primer Buffer and then added to the reagent cartridge. Make sure that the reagent cartridge is thawed and inspected before proceeding.
| 1. | If frozen, thaw each custom primer to be used. |
| 2. | Use NovaSeq X Series Custom Primer Buffer to dilute custom primers to yield the following volumes at 0.3 µM final concentration. When combining primers to make the custom Index Read or any custom read primer mixture, each primer should be at 0.3 µM final concentration. |
|
Custom Primer |
Volume (ml) for 1.5B Flow Cell |
Volume (ml) for 10B or 25B Flow Cell |
|---|---|---|
|
Custom Read 1 primer |
3.0 |
5.0 |
|
Custom Read 2 primer |
3.0 |
5.0 |
|
Custom Index 1 and 2 primer mix |
5.0 |
6.0 |
| 1. | Use a polynit heatseal wipe to wipe clean the foil seal covering each custom primer position. |
|
Position |
Custom Primer |
|---|---|
|
CP1 |
Custom Read 1 primer |
|
CP2 |
Custom Read 2 primer |
|
CP3 |
Custom Index 1 and 2 primer mix |
| 2. | Using a clean pipette tip, pierce the foil seal covering each custom primer position. |
| 3. | Add the following volumes of custom primer to the appropriate position on the reagent cartridge. Avoid touching the foil seal as you dispense the primer. |
|
Position |
Custom Primer |
Volume (ml) for 1.5B Flow Cell |
Volume (ml) for 10B or 25B Flow Cell |
|---|---|---|---|
|
CP1 |
Custom Read 1 primer |
3.0 |
5.0 |
|
CP2 |
Custom Read 2 primer |
3.0 |
5.0 |
|
CP3 |
Custom Index 1 and 2 primer mix |
5.0 |
6.0 |