Prepare Wash Solution and NaOCl

1. Add 400 ml laboratory-grade water to a 500 ml centrifuge bottle.
2. Add 200 µl 100% Tween 20 or an analogous volume of a diluted tween solution. For example, 1 ml 20% Tween 20 in laboratory-grade water.

Use a freshly prepared dilution of Tween 20 to limit the introduction of contaminants into the fluidics system.

3. Invert to mix.
4. Combine the following volumes in a 50 ml centrifuge tube to prepare 49 ml of 0.12% reagent grade NaOCl:
5% reagent grade NaOCl (1.2 ml)
Deionized water (48 ml)

Only use reagent grade NaOCl. Avoid general-purpose bleach products as these can contain ammonia compounds, which might lead to runs with low percentage passing filter reads.

5. Invert to mix.