Start a Manual Run
Use the following instructions to initiate a manual run. Depending on the run type, a sample sheet might be required or optional.
|
1.
|
If you are not signed in, follow the instructions in Sign In and Sign Out. |
|
2.
|
Navigate to the Start screen, and then select Start. |
|
4.
|
On the NovaSeq X Plus, select an instrument side for a single-sided run or select both sides for a dual-sided run. |
|
5.
|
At the top of the runs list, select Manual. |
Planned is the default run selection. To change the default setting, refer to Customize System Settings.
|
6.
|
Select from the following run types: |
|
•
|
Manual run (generate BCLs only) |
|
•
|
Import sample sheet for local analysis |
|
7.
|
For manual runs that generate BCLs only, configure your run as follows. |
|
b.
|
The run name can contain a maximum of 225 alphanumeric characters, spaces, dashes, and underscores. |
|
c.
|
Select single or paired-end for the read type. |
|
d.
|
Enter the number of cycles performed in each read and the length of each index read. |
|
e.
|
The total number of read cycles and index cycles cannot exceed the number of cycles specified by the reagent kit. |
|
•
|
Read 1—Enter the number of cycles for Read 1. |
|
•
|
Index 1—Enter the length of the index read for Index 1. For a PhiX-only run, enter 0 in both index fields. |
|
•
|
Index 2—Enter the length of the index read for Index 2. |
|
•
|
Read 2—Enter the number of cycles for Read 2. This value is typically the same as the Read 1 value. |
|
f.
|
[Optional] Select your sample sheet. |
|
8.
|
For manual runs with local secondary analysis, the sample sheet is required. Select Select, and then open the sample sheet v2 file. |
|
9.
|
Select Review, and then review the run information. |
|
10.
|
To use custom primers, select the appropriate Custom primer checkboxes. |
For more information on custom primers, refer to Custom Primers.
|
11.
|
If the instrument is connected to the cloud, confirm the cloud run setting. |
|
12.
|
Confirm the output folder. |
For runs with cloud storage enabled, the following options are available:
|
•
|
To transfer run data to a local storage folder in addition to the cloud, provide an output folder. |
|
•
|
To only transfer run data to the cloud, select Don't transfer run data to external output folder. |
To change the default output folder, refer to Specify the Default Output Folder Location for more information.
|
13.
|
If performing local secondary analysis, confirm BCL data folder transfer setting. |
To update the default run data management selection, refer to Customize System Settings.
|
14.
|
[Optional] Select a custom recipe file. |
If using the Illumina Stranded Total RNA Prep with Ribo-Zero Plus kit or the Illumina Stranded mRNA Prep kit, selecting the custom recipe is required. Refer to Dark Cycle Sequencing for more information.
|
15.
|
When finished, select Load consumables. |