Laboratory Requirements
Use the specifications and requirements provided in this section to set up your lab space.
Position the instrument to allow proper ventilation and access for servicing. Use the following minimum clearance dimensions to make sure that the instrument is accessible from all sides.
|
Access |
Minimum Clearance |
|---|---|
|
Sides |
Allow at least 30 cm (12 in) on each side of the instrument. |
|
Rear |
Allow at least 15.25 cm (6 in) behind the instrument. |
|
Top |
Allow at least 30 cm (12 in) above the instrument (lowered monitor). |
| • | Make sure that you can reach around the right side of the instrument to access the power switch on the back panel. |
| • | Position the instrument so that you can quickly disconnect the power cord from the outlet. |
The instrument includes precision optical elements. Place the instrument on a sturdy and level lab bench away from sources of vibration.
During sequencing runs, use the following best practices to minimize continuous and intermittent vibrations and ensure optimal performance.
| • | Keep the lab bench free of potential vibration sources, such as: |
| – | Shakers, vortex mixers, centrifuges, drawers, cabinets, and shelves that can cause unintentional shocks to the bench surface. |
| – | Pressurized air or nitrogen and other major air flows. |
| • | Keep the clearance area free of clutter. |
| • | Do not place used consumables or other lab tools and accessories on the instrument. |
| • | When interacting with the instrument, use only the touch screen monitor and the recommended workflow for loading and unloading consumables. |
| • | Do not impact instrument surfaces. |
When you receive iSeq 100 i1 Reagent kits, promptly store components at the appropriate conditions to ensure proper performance:
| • | Store at the indicated temperatures. |
| • | Do not open the white foil packaging until instructed to do so. The cartridge is thawed in the bag. |
| • | Position the cartridge so that the package label faces up. |
| • | Store the cartridge for at least one day before thawing in a water bath. |
The following table provides storage temperature and dimensions for components included in iSeq 100 i1 Reagent v2 kits, a single-use reagent kit.
|
Component |
Storage Temperature |
Length |
Width |
Height |
|---|---|---|---|---|
|
Cartridge |
-25°C to -15°C |
19.6 cm (7.7 in) |
13.7 cm (5.4 in) |
13 cm (5 in) |
|
Flow cell |
2°C to 8°C* |
10.2 cm (4 in) |
10.2 cm (4 in) |
2.5 cm (1 in) |
*Shipped at room temperature.