General Safety Warnings
Make sure that all personnel are trained in the correct operation of the
|
Follow all operating instructions when working in areas marked with
this label to minimize risk to personnel or the |
Electrical Safety Warnings
Do not remove the outer panels from the instrument. There are no user-serviceable components inside. Operating the instrument with any of the panels removed creates potential exposure to line voltage and DC voltages.
|
The instrument is powered by 100–240 volts AC operating at 50/60 Hz. Hazardous voltage sources are located behind the rear and side panel, but can be accessible if other panels are removed. Some voltage is present on the instrument even when the instrument is turned off. Operate the instrument with all panels intact to avoid electrical shock. |
|
Type |
Specification |
|---|---|
|
Line Voltage |
100–240 VAC at 50/60 Hz |
|
Peak Power Consumption |
80 Watts |
An electrical ground is required. If the voltage fluctuates more than 10%, a power line regulator is required.
Position the instrument so that you can quickly disconnect the power cord from the outlet.
|
|
The instrument has a connection to protective earth through the enclosure. The safety ground on the power cord returns protective earth to a safe reference. The protective earth connection on the power cord must be in good working condition when using this device. |
The power entry module includes two input fuses on the high-voltage input lines. These fuses are size 5 mm × 20 mm and rated for 10 Amps, 250 VAC, Slow Blow.
Do not operate the instrument with any of the panels removed.