Microseal 'B' adhesive seals |
|
1.
|
Prepare the following consumables: |
|
•
|
ATL4—Vortex to mix, and then centrifuge briefly. Make sure that ATL4 is dissolved completely. |
|
2.
|
Save the following ATL program on the thermal cycler: |
|
•
|
Choose the preheat lid option and set to 100°C |
|
•
|
Set the reaction volume to 90 μl |
|
1.
|
Add 35 µl ATL4 to each well of the LP plate. |
|
2.
|
Apply Microseal 'B' and shake at 1800 rpm for 1 minute. |
|
4.
|
Place on the preprogrammed thermal cycler and run the ATL program. |
|
5.
|
When the program reaches the 4°C hold, immediately proceed to the next step. |