Microseal 'B' adhesive seal |
|
1.
|
Prepare the following consumables: |
|
•
|
ERP6—Vortex to mix, and then centrifuge briefly. |
|
2.
|
Save the following ERP program on the thermal cycler: |
|
•
|
Choose the preheat lid option and set to 100°C |
|
•
|
Set the reaction volume to 55 µl |
|
1.
|
Remove the LP plate from ice. |
|
2.
|
Add 10 µl ERP6 to each well of the LP plate. |
|
3.
|
Apply Microseal 'B' and shake at 1800 rpm for 1 minute. |
|
5.
|
Place on the preprogrammed thermal cycler and run the ERP program.
|
|
6.
|
When the program reaches the 4°C hold, immediately proceed to the next step. |