MRR (Methylation Resuspension Reagent) |
|
•
|
UMI3 (Unique Molecular Identifier) |
|
•
|
Microseal 'B' adhesive seals |
|
1.
|
Prepare the following consumables: |
|
•
|
LIGX—Before use, centrifuge briefly, pipette to mix, and then centrifuge briefly again. |
|
•
|
MRR—Bring to room temperature to prepare for Clean Up Ligation. |
|
•
|
UMI3—Vortex on medium-high speed for at least 10 seconds to mix, and then centrifuge briefly. |
|
2.
|
Save the following LIG program on the thermal cycler: |
|
•
|
Choose the preheat lid option and set to 42°C |
|
•
|
Set the reaction volume to 100 μl |
|
1.
|
Add 5 μl LIGX to each well of the LP plate. |
|
2.
|
Add 5 μl UMI3 adapters to each well. |
|
3.
|
Apply Microseal 'B' and shake at 1800 rpm for 1 minute. |
|
5.
|
Place on the preprogrammed thermal cycler and run the LIG program. |
|
6.
|
When the program reaches the 4°C hold, immediately proceed to the next step. |