Fragment FFPE DNA
This snippet causes dropdowns to open on page load. This text/paragraph does not appear in the output.
This process fragments FFPE DNA to a 100–350 bp fragment size using the Covaris LE220-plus, ME220 and R230 focused-ultrasonicator. Covaris shearing generates dsDNA fragments with 3’ and 5’ overhangs.
| • | TEB (TE Buffer, user-supplied) |
| • | Covaris 8 microTUBE strip with foil seals (if using LE220-plus Covaris instrument). |
| • | 96-well PCR plate, semiskirted |
| • | Covaris 8 microTUBE strip with sample rack and waveguide (if using ME220 Covaris instrument). |
| • | 96 microTUBE plate with SonoLab 10.0.1 (or higher), black (PN 520078 if using R230 Covaris instrument) |
| 1. | Turn on and set up the Covaris instrument according to manufacturer guidelines. Some Covaris instruments require ~1 hour to de-gas. |
| • | On the Covaris LE220-plus and Covaris R230, select the correct shearing protocol and appropriate columns. |
| 2. | Make sure that the instrument contains correct water levels before proceeding. |
| 3. | Thaw FFPE DNA samples on ice. |
| 4. | Pipette 10 times to mix. |
| 5. | Centrifuge tube briefly to collect droplets. |
| 6. | Vortex TEB to mix. |
| 7. | Label a new 96-well semiskirted PCR plate LP. |
| 1. | Dilute each FFPE DNA sample with TEB as follows: |
| • | For 40 ng input, prepare 53 μl at 0.89 ng/µl. The assay requires a minimum extraction concentration of 3.93 ng/μl to allow for at least 41 µl TEB of the 53 µl volume. |
| 2. | Pipette each diluted FFPE DNA sample 10 times to mix. |
| 3. | Transfer 52 μl of each diluted FFPE DNA sample into a separate well of the ultrasonicator tube. Load the FFPE DNA into Covaris tube slowly to avoid creating bubbles. |
| 4. | Fill any unused Covaris 8 microTUBE Strip wells with 52 μl TEB. |
| 5. | Seal the microtube Strip with the foil seal. |
| 6. | Centrifuge briefly. |
| 7. | If using the Covaris LE220-plus, ME220 or R230 model, fragment the DNA using the following settings. |
|
Setting |
LE220-plus |
ME220 |
R230 |
|---|---|---|---|
|
Peak incident power |
450 watts |
50 watts |
450 watts |
|
Duty Factor |
30% |
30% |
25% |
|
Cycles per burst |
200 |
1000 |
600 |
|
Treatment time |
250 seconds |
10 seconds |
— |
|
Temperature |
7°C |
12°C |
10°C |
|
Pulse repeats |
— |
20 |
— |
|
Average power |
— |
15 watts |
— |
|
Other |
— |
Wave guide |
— |
|
Iterations |
— |
— |
20 |
|
Repeat Process Treatment Duration |
— |
— |
10 seconds |
|
Delay Duration |
— |
— |
10 seconds |
|
Dithering |
— |
— |
1.5 mm @ 10 mm/s |
| 8. | After sonication, centrifuge Covaris 8 microTUBE Strip briefly to collect droplets. |
| 9. | Transfer 45 µl each sheared FFPE DNA sample to the corresponding well of the LP plate. |
SAFE STOPPING POINT
If you are stopping, apply Microseal 'B' adhesive seal to the LP plate and briefly centrifuge at 280 × g. Store at -25°C to -15°C for up to 7 days.