Clean Up Libraries
This snippet causes dropdowns to open on page load. This text/paragraph does not appear in the output.
This step uses double-sided bead purification procedure to purify the amplified and indexed libraries.
Consumables
|
•
|
IPB (Illumina Purification Beads) |
|
•
|
RSB (Resuspension Buffer) |
|
•
|
EtOH (Freshly prepared 80% ethanol) |
|
•
|
96-well 0.8 ml Polypropylene Deepwell Storage Plate (MIDI plate) (2) |
|
•
|
Microseal 'B' adhesive seal |
|
•
|
Microseal 'F' foil seal |
About Reagents
|
–
|
Must be at room temperature before use. |
|
–
|
Vortex to resuspend before each use. |
|
–
|
Aspirate and dispense slowly due to the viscosity of the solution. |
Preparation
|
2.
|
For each sample, prepare 400 µl fresh 80% EtOH from absolute ethanol. Including an overage of 20% is recommended. |
Procedure
|
1.
|
Centrifuge at 280 × g for 1 minute to collect contents at the bottom of the well. |
|
2.
|
Place the plate on the magnetic stand and wait until the liquid is clear (~5 minutes). |
|
3.
|
Transfer 45 µl supernatant from each well of the PCR plate to the corresponding well of a new MIDI plate. |
|
4.
|
Resuspend IPB as follows. |
|
a.
|
To mix, invert the bottle manually for 2 minutes, at a rate of 1 inversion per second. While inverting, rotate the bottle 90 degrees every 30 seconds. |
|
b.
|
If beads are still adhered to the walls of the container, invert the bottle manually for an additional 1 minute. |
|
5.
|
For standard DNA input > 500 bp gDNA, blood, or saliva, perform the following steps. |
|
a.
|
Add 40 µl nuclease-free water to each well-containing supernatant. |
|
b.
|
Add 45 µl IPB to each well-containing supernatant. |
|
c.
|
Pipette each well 10 times to mix. Alternatively, seal the plate and use a plate shaker at 1600 rpm for 1 minute. |
|
d.
|
Seal the plate and incubate at room temperature for 5 minutes. |
|
e.
|
Place on the magnetic stand and wait until the liquid is clear (~5 minutes). |
|
f.
|
During incubation, thoroughly vortex and resuspend the IPB, and then add 15 µl to each well of a new MIDI plate. |
|
g.
|
Transfer 125 µl supernatant from each well of the first plate into the corresponding well of the new MIDI plate containing 15 µl undiluted IPB. |
|
h.
|
Pipette each well in the MIDI plate 10 times to mix. Alternatively, seal the plate and use a plate shaker at 1600 rpm for 1 minute. |
|
i.
|
Discard the first plate. |
|
6.
|
For small PCR amplicon input < 500 bp, perform the following steps. |
|
a.
|
Add 81 µl IPB to each well of the MIDI plate containing supernatant. |
|
b.
|
Pipette each well 10 times to mix. Alternatively, seal the plate and use a plate shaker at 1600 rpm for 1 minute. |
|
7.
|
Incubate the sealed MIDI plate at room temperature for 5 minutes. |
|
8.
|
Place on the magnetic stand and wait until the liquid is clear (~5 minutes). |
|
9.
|
Without disturbing the beads, remove and discard supernatant. |
|
10.
|
Wash beads as follows. |
|
a.
|
With the plate on the magnetic stand, add 200 µl fresh 80% EtOH without mixing. |
|
b.
|
Incubate for 30 seconds. |
|
c.
|
Without disturbing the beads, remove and discard supernatant. |
|
11.
|
Wash beads a second time. |
|
12.
|
Use a 20 µl pipette to remove and discard residual EtOH. |
|
13.
|
Air-dry on the magnetic stand for 5 minutes. |
|
14.
|
Remove from the magnetic stand. |
|
15.
|
Add 32 µl RSB to the beads. |
|
16.
|
Pipette to resuspend.
|
|
17.
|
Incubate at room temperature for 2 minutes. |
|
18.
|
Place the plate on the magnetic stand and wait until the liquid is clear (~2 minutes). |
|
19.
|
Transfer 30 µl supernatant to a new 96-well PCR plate. |
SAFE STOPPING POINT
If you are stopping, seal the plate with Microseal 'B' adhesive seal or Microseal 'F' foil seal, and store at ‑25°C to ‑15°C for up to 30 days.