Tagment Genomic DNA
This snippet causes dropdowns to open on page load. This text/paragraph does not appear in the output.
This step uses the Bead-Linked Transposomes (BLT) to tagment DNA, which is a process that fragments and tags the DNA with adapter sequences.
| • | BLT (Bead-Linked Transposomes) |
| • | TB1 (Tagmentation Buffer 1) |
| • | Nuclease-free water |
| • | 8-tube strip |
| • | 96-well PCR plate |
| • | Microseal 'B' adhesive seal |
This set of reagents contains potentially hazardous chemicals. Personal injury can occur through inhalation, ingestion, skin contact, and eye contact. Wear protective equipment, including eye protection, gloves, and laboratory coat appropriate for risk of exposure. Handle used reagents as chemical waste and discard in accordance with applicable regional, national, and local laws and regulations. For additional environmental, health, and safety information, refer to the SDS at support.illumina.com/sds.html.
| • | BLT |
| – | Must be stored upright so that the beads are always submerged in the buffer. |
| – | Do not use BLT that has been stored below 2°C. |
| 1. | Prepare the following consumables: |
| • | BLT—Vortex to mix. Do not centrifuge before pipetting. |
| • | TB1—Vortex to mix. |
| 2. | Save the following TAG program on the thermal cycler: |
| • | Choose the preheat lid option and set to 100°C |
| • | Set the reaction volume to 50 µl |
| • | 55°C for |
| • | Hold at 10°C |
| 1. | Add 2–30 µl DNA to each well of a 96-well PCR plate so that the total input amount is |
| 2. | If DNA volume is < 30 µl, add nuclease-free water to the DNA samples to bring the total volume to 30 µl. |
| 3. | Vortex BLT for 10 seconds to resuspend. Repeat as necessary. |
| 4. | For each sample, combine the following volumes to prepare the Tagmentation Master Mix. Multiply each volume by the number of samples being processed. |
| • | BLT |
| • | TB1 |
These volumes produce
| 5. | Vortex the Tagmentation Master Mix for 10 seconds to resuspend. |
| 6. | Divide the Tagmentation Master Mix volume equally into an 8-tube strip. |
| 7. | Using a multichannel pipette, transfer 20 µl Tagmentation Master Mix from the 8-tube strip to each well of the plate containing a sample. |
Use fresh tips for each sample column.
| 8. | Discard the 8-tube strip after the Tagmentation Master Mix has been dispensed. |
| 9. | Pipette each sample 10 times to resuspend. |
| 10. | Place on the preprogrammed thermal cycler and run the TAG program. |