Fragmentation, End Repair, & A-Tailing
This snippet causes dropdowns to open on page load. This text/paragraph does not appear in the output.

• | LPB (Library Prep Buffer) |
• | LPE (Library Prep Enzymes) |

1. | Prepare the following consumables: |
• | LPB—Vortex to mix, and then centrifuge briefly. Keep on ice. |
• | LPE—Pipette to mix, and then centrifuge briefly. Keep on ice. |
2. | Save the following cDNA Fragmentation program on the thermal cycler: |
• | Set the lid temperature to 105°C |
• | Hold at 4°C |
• | 30°C for 6 minutes |
• | 65°C for 30 minutes |
• | Hold at 4°C |
3. | Run the cDNA Fragmentation program to prechill the thermal cycler block to 4°C and preheat the lid to 105°C. |

1. | Combine the following volumes to prepare Library Prep Master Mix. |
Reagent |
Volume Per Reaction (μl) |
4.4X Reaction Master Mix (µl) |
8.8X Reaction Master Mix (µl) |
---|---|---|---|
LPB |
4 |
17.6 |
35.2 |
LPE |
6 |
26.4 |
52.8 |
Total |
10 |
44 |
88 |
2. | Vortex to mix, and then centrifuge briefly. |
3. | Add 10 μl Library Prep Master Mix to each tube containing 40 μl of cDNA. |
The total volume is 50 μl.
4. | Vortex to mix or pipette mix 10 times at the 25 µl stroke. |
Make sure to mix completely.
5. | Centrifuge briefly. Place the samples on ice. |
6. | Place on the prechilled thermal cycler, and then use the skip step function to run the next steps of the cDNA Fragmentation program. |
7. | When the thermal cycler program reaches 4°C, immediately remove the sample and proceed to the next step. |