Prepare for Protocol

This preparation is required to perform the protocol steps leading up to the next safe stopping point.

1. Remove sample from storage.
2. Remove third-party reagents from storage.
Third-Party Reagents

Reagent

Storage

Thaw Instructions

RNase Inhibitor*

Varies for third‑party reagents

Keep on ice.

Refer to the manufacturer instructions as needed.

[Cryopreserved cells] Thawing media

Varies for third‑party reagents

Refer to the manufacturer instructions as needed.

* For cell and nuclei suspensions, use user-supplied RNase Inhibitor.

3. Remove the reagents from the box and thaw as follows.
15°C to 30°C Storage

Reagent

Box Name

Thaw Instructions

PAR

Illumina Single Cell 3′ RNA Capture, T100, Box 1

Use at room temperature.

CLB3

Illumina Single Cell 3′ RNA Capture, T100, Box 1

Use at room temperature.

-25°C to -15°C Storage

Reagent

Box Name

Thaw Instructions

[Cell preparation] CSB

Illumina Single Cell 3′ RNA Capture, T100, Box 3

For cryopreserved cells—Warm to 37°C in a water bath.

For fresh cells—Keep cold.

[Nuclei preparation] NSB

Illumina Single Cell 3′ RNA Capture, T100, Box 3

Use at room temperature.

RNI*

Illumina Single Cell 3′ RNA Capture, T100, Box 3

Keep on ice.

25% BSA

Illumina Single Cell 3′ RNA Capture, T100, Box 3

Thaw on ice.

* For capture and lysis, use Illumina-supplied RNI.

-85°C to -65°C Storage

Reagent

Box Name

Thaw Instructions

T100 PIP

Illumina Single Cell 3′ RNA Capture, T100, Box 4

Thaw one tube per sample on ice. Stable on ice for up to 5 hours.